November 30th 2023
Important Safety Notice: ISC UltraLink
Safety Notice/Product Recall Following on from our inspection notice issued 17 November 2023, our team has been intensively investigating the potential risk and the root cause of the problem. Our
investigation identified that the press fixture used to secure the assembly of the SafeLock button moved outside of specification, creating the potential for a faulty press fit of the button, which can lead to
the button turning open and coming apart.
STOP USE
Due to the risk of unintentional
opening of the device, which could result in a life-threatening situation, we are issuing this notice instructing users to stop using their device, if it falls within the range of serial numbers:
LK100A1.2: 170431—177209; or LK101A1.2: 170433—176974
Manufacturing corrective actions have been introduced to repair and ensure that the press fixture remains
within specification; and additional procedures have been introduced to ensure that the SafeLock buttons are secured correctly. Validation
testing has been successfully completed and production of UltraLinks has resumed, with this new process in place.
To start the recall process please register here